Skip to main content

Phantom-based acquisition time and image reconstruction parameter optimisation for oncologic FDG PET/CT examinations using a digital system

Abstract

Background

New-generation silicon-photomultiplier (SiPM)-based PET/CT systems exhibit an improved lesion detectability and image quality due to a higher detector sensitivity. Consequently, the acquisition time can be reduced while maintaining diagnostic quality. The aim of this study was to determine the lowest 18F-FDG PET acquisition time without loss of diagnostic information and to optimise image reconstruction parameters (image reconstruction algorithm, number of iterations, voxel size, Gaussian filter) by phantom imaging. Moreover, patient data are evaluated to confirm the phantom results.

Methods

Three phantoms were used: a soft-tissue tumour phantom, a bone-lung tumour phantom, and a resolution phantom. Phantom conditions (lesion sizes from 6.5 mm to 28.8 mm in diameter, lesion activity concentration of 15 kBq/mL, and signal-to-background ratio of 5:1) were derived from patient data. PET data were acquired on an SiPM-based Biograph Vision PET/CT system for 10 min in list-mode format and resampled into time frames from 30 to 300 s in 30-s increments to simulate different acquisition times. Different image reconstructions with varying iterations, voxel sizes, and Gaussian filters were probed. Contrast-to-noise-ratio (CNR), maximum, and peak signal were evaluated using the 10-min acquisition time image as reference. A threshold CNR value ≥ 5 and a maximum (peak) deviation of ± 20% were considered acceptable. 20 patient data sets were evaluated regarding lesion quantification as well as agreement and correlation between reduced and full acquisition time standard uptake values (assessed by Pearson correlation coefficient, intraclass correlation coefficient, Bland–Altman analyses, and Krippendorff’s alpha).

Results

An acquisition time of 60 s per bed position yielded acceptable detectability and quantification results for clinically relevant phantom lesions ≥ 9.7 mm in diameter using OSEM-TOF or OSEM-TOF+PSF image reconstruction, a 4-mm Gaussian filter, and a 1.65 × 1.65 x 2.00-mm3 or 3.30 × 3.30 x 3.00-mm3 voxel size. Correlation and agreement of patient lesion quantification between full and reduced acquisition times were excellent.

Conclusion

A threefold reduction in acquisition time is possible. Patients might benefit from more comfortable examinations or reduced radiation exposure, if instead of the acquisition time the applied activity is reduced.

Peer Review reports

Background

New-generation “digital” positron emission tomography/computed tomography (PET/CT) systems show a higher spatial and coincidence time resolution than previous-generation systems mainly because they use silicon-photomultipliers (SiPMs) that exhibit a higher detector sensitivity than previously used photomultiplier-tubes (PMTs) [1, 2]. Consequently, lesion detectability and image quality are improved [3,4,5]. Therefore, the acquisition time can be reduced while maintaining diagnostic image quality [6, 7].

Advantages of a reduced acquisition time include improved cost effectiveness and patient comfort, especially for pain-stricken, dyspnoeic, or paediatric patients [8, 9]. Moreover, motion artifacts can be reduced [10]. A typical clinical indication for 18F-FDG PET/CT scans is interim and final staging during chemotherapy of lymphoma patients [11]. As solid lymphomas belong to the most frequent malignancies in children, they are a common indication for paediatric PET scans [12, 13].

Alternatively, the amount of administered activity could be reduced, since, in a first approximation, a linear correlation between acquisition time and administered activity can be assumed [7, 14]. Thus, the radiation exposure for medical staff and patients could be reduced. Typical young lymphoma patients who respond to chemotherapy and show a high long-term survival could benefit from a reduced risk for secondary radiation-induced malignancy [15].

According to the current EANM guidelines for oncologic 18F-FDG PET/CT imaging [16] a typical acquisition time is 3 min per bed position, if activities of 2–3 MBq per kg patient weight are administered. We hypothesise that by application of a current-generation SiPM-based Biograph Vision PET/CT systems the administered activity can be reduced while maintaining lesion detectability and image quantification. A preliminary patient evaluation by our group indicated that a reduction by a factor of one-third is feasible while maintaining diagnostic quality [6]. However, comprehensive phantom measurements to confirm the hypothesis under defined conditions are, to the best of the authors’ knowledge, yet missing.

The aim of this study was to optimise the 18F-FDG PET acquisition time and image reconstruction parameters (image reconstruction algorithm, number of iterations, voxel size, Gaussian filter) derived from imaging of several phantoms, whose setup and preparation were based on clinical data of lymphoma patients. Moreover, previously acquired patient data are re-evaluated to validate the phantom results.

Methods

Phantom setup and preparation

To simulate nodal and extra-nodal lymphoma lesions, a soft-tissue tumour phantom and a bone-lung tumour phantom were used. Thorax and skeletal system are, among others, common localisations of extra-nodal lymphoma manifestations [17, 18]. Additionally, a line phantom was used for determination of spatial resolution.

To derive phantom conditions that mimic clinically realistic conditions, 16 randomly selected FDG PET/CT scans of clinical routine lymphoma patients were evaluated. For comparability, measured activity concentrations were normalised to patient weight and administered activity (assuming a 70-kg patient mass and an administered activity of 3 MBq/kg as recommended by current guidelines [16]); signal-to-background-ratios (SBRs) and volumes were estimated for a total of 31 lesions. Background activity concentrations were evaluated in horse-shoe-shaped volumes-of-interest (VOIs) surrounding the respective lesion; volumes were determined using a 3D-isocontour approach with a 50%-of-maximum threshold [16] or using CT data. 24/31 lesions were nodal lesions, the remaining 7/31 were bone lesions. Mean (minimum–maximum) activity concentration and volume were 12.2 (4.2–43.8) kBq/mL and 56.7 (0.7–328.1) mL, respectively. The mean (minimum–maximum) SBR was 4.0 (2.1–4.6). Based on these clinical data, a representative activity concentration of 15 kBq/mL and a SBR of 5:1 were selected. Since detectability and quantification is most challenging for small lesions, lesion volumes reflecting the lower range of volumes in the clinical evaluation were selected.

Soft-tissue tumour phantom

The soft-tissue phantom consisting of a torso-shaped NEMA phantom (Data Spectrum Corporation, Durham, USA) that contains 6 small spheres (Fig. 1A) was designed to simulate hot lesions in a uniform warm background. Both spheres and phantom cavity were filled with radioactive solution. The largest sphere in the original NEMA phantom (37.0-mm diameter) was replaced by a small sphere (6.5-mm diameter). Thus, the resulting sphere diameters ranged from 6.5 mm to 28.0 mm (Fig. 1A). Each sphere was filled with 18F-FDG in aqueous solution. Of note, the second smallest sphere (9.7-mm diameter) represents a typical clinical threshold size for conspicuous lymph nodes [19].

Fig. 1
figure 1

Frontal and lateral views (central plane) of a schematic representation of the soft-tissue (A), the bone-lung tumour phantom (B), and the resolution phantom (C)

Bone-lung tumour phantom

The bone-lung tumour phantom consisting of three cylindrical tubes inside the torso-shaped NEMA phantom (Fig. 1B) was designed to simulate hot pulmonary or osseous tumour lesions in a non-radioactive environment. To represent the CT density of lung-tissue (–800 HU), one tube was filled with cork powder (3 × 10–4 g/mm3-density, grain size 0.0–0.2 mm; Dumke Group, Rothenstein, Germany). To simulate the CT density of different kinds of bone lesions, two tubes were filled with dipotassium hydrogen phosphate (K2HPO4) solution in different concentrations. The calculated mass absorption coefficient for K2HPO4 at 500 keV is 0.0868 cm2/g [20] closely resembling the mass absorption coefficient for cortical bone of 0.0902 cm2/g [20]. Therefore, K2HPO4 is commonly used to simulate bone lesions in phantom investigations of attenuation correction in nuclear medicine imaging techniques [21, 22]. In the bone-lung tumour phantom, the resulting CT densities of 500 HU and 1000 HU reflect the spherical head of the humerus (538 HU) and the femoral shaft (1239 HU) [23]. Each tube contained two spheres (8.5-mm and 19.4-mm diameter) filled with 18F-FDG in aqueous solution. Spheres and phantom cavity were filled with radioactive solution. The tubes, however, contained no radioactive solution.

Resolution phantom

The resolution phantom contains multiple line sources consisting of polyethylene tubing (that are mounted into the abdominal torso phantom in orthogonal orientation to the transverse plane, Fig. 1C). The inner diameter of the line sources is 0.5 mm and they are looped back through the phantom to provide distances of 10 mm and 100 mm from the central axis of the field-of-view. They were filled with 18F-FDG in aqueous solution with an activity concentration of 7.1 MBq/mL; the total activity in the scanner was 21 MBq. For spatial resolution quantification with scatter, the phantom cavity was filled with non-radioactive water.

PET data acquisition and image reconstruction

All phantom PET data were acquired on a SiPM-based Biograph Vision 600 PET/CT scanner (Siemens Healthineers, Erlangen, Germany) using a single bed position in list-mode. Detailed scanner specifications are given in Table 1. For attenuation correction, additional CT data were acquired (acquisition conditions: Care DOSE 4D, quality reference 160 mAs; CARE kV, quality reference 120 kV).

Table 1 Detailed specifications of the Biograph Vision PET/CT system. LSO Lutetium Oxyorthosilicate

Soft-tissue and bone-lung tumour phantoms

PET data were acquired for 10 min in list-mode (one single bed position) and resampled into time frames from 30 to 300 s in 30-s increment to simulate different acquisition times. The 10-min acquisition time data were used as reference standard.

PET images were reconstructed using three-dimensional ordinary Poisson ordered-subset expectation maximization (OSEM) with or without time-of-flight option (TOF) and with or without point-spread-function modelling (PSF) resulting in four different combinations of reconstruction algorithms (i.e., OSEM, OSEM-TOF, OSEM-PSF, OSEM-TOF+PSF). Moreover, different numbers of iterations (4, 6, and 8 for TOF-based and 10,12, and 14 for non-TOF reconstruction algorithms) were evaluated. For both non-TOF and TOF-based reconstruction groups, the smallest investigated number of iterations reflects the manufacturer’s recommendation for the respective reconstruction algorithm. Of note, the number of subsets was 5 and was fixed in all image reconstructions.

The applied matrices (220 or 440) and Gaussian filters (2 mm or 4 mm) resulted in 4 different combinations of voxel sizes and Gaussian filters (i.e., 1.65 × 1.65 x 2.00-mm3 voxel size and 2-mm Gaussian filter, 3.30 × 3.30 x 3.00-mm3 voxel size and 2-mm Gaussian filter, 1.65 × 1.65 x 2.00-mm3 voxel size and 4-mm Gaussian filter, 3.30 × 3.30 x 3.00-mm3 voxel size and 4-mm Gaussian filter).

Resolution phantom

To determine the spatial resolution at high counting statistics, the acquisition time was increased to 30 min. PET data were reconstructed using OSEM-TOF with for 4 iterations and OSEM-TOF+PSF with 4 iterations only. Two different spatial resolutions were defined. For estimation of the system spatial resolution, an 880 matrix was applied (resulting voxel size of 0.83 × 0.83 x 2.00 mm3) without Gaussian filter. The clinical spatial resolution further includes Gaussian filtering (2-mm or 4-mm filter level) and was calculated from the system spatial resolution as the convolution of two Gaussian functions [24] (see Eq. 4).

Analyses of phantom images

Soft-tissue and bone-lung tumour phantom

Three activity concentration values were determined for the spherical objects and one activity concentration for the background. Sphere average activity concentrations were determined in spherical VOIs whose diameter matched that of the real spheres. Sphere average activity concentrations were applied (see below) to calculate the contrast-to-noise ratio (CNR). Maximum activity concentrations were derived within the sphere boundary; peak activity concentrations were ascertained in peak VOIs (sphere of 1 mL-volume) that were placed at the centre of the respective spherical phantom insert. Peak and maximum activity concentrations were used to assess the activity concentration accuracy at different acquisition times (see below). For the soft-tissue tumour phantom only, image noise was evaluated. For this purpose, the average activity concentration and its standard deviation were determined in a background VOI. The background VOI consisted of 60 two-dimensional 37-mm diameter circular background regions-of-interest that were placed in the central plane of the spherical phantom inserts and the two adjacent planes in each direction (12 regions-of-interest in each plane) according to NEMA "Standard for Performance Measurements of Positron Emission Tomographs NU 2–2012″ [25].

To evaluate the lesion detectability, the contrast-to-noise ratio (CNR) was calculated for the spherical inserts in the soft-tissue tumour phantom as previously described [26]:

$$\begin{array}{c}CNR = \frac{{C}_{\mathrm{avg}}\ - \ {C}_{\mathrm{bgr}}}{{SD}_{\mathrm{bgr}}}\end{array}$$
(1)

using the average sphere activity concentration (denoted by Cavg), the average activity concentration in the background VOI (denoted by Cbgr), and the standard deviation of the activity concentration in the background VOI (denoted by SDbgr).

For both the soft-tissue tumour and the bone-lung tumour phantom, the accuracy of activity concentration measurements in short-acquisition time images was evaluated. For this purpose, maximum and peak activity concentration ratios ACRmax and ACRpeak were calculated using the 10-min scan as reference:

$$\begin{array}{c}{\mathrm{ACR}}_{\mathrm{max}} = \frac{{C}_{\mathrm{max}}}{{C}_{\mathrm{max},\mathrm{ref}}}\end{array}$$
(2)

and

$$\begin{array}{c}{\mathrm{ACR}}_{\mathrm{peak}} = \frac{{C}_{\mathrm{peak}}}{{C}_{\mathrm{peak},\mathrm{ref}}}\end{array}$$
(3)

defining the maximum (peak) sphere activity concentration, Cmax (Cpeak), at varying acquisition times and the respective maximum (peak) activity concentration in the 10-min reference image Cmax,ref (Cpeak,ref).

Of note, maximum and peak activity concentrations can be regarded as surrogates of the clinically established maximum and peak standardised uptake values SUVmax and SUVpeak. SUVmax and SUVpeak are used for patient PET data to describe activity concentration measurements, which are normalised to patient mass and administered activity. Therefore, they cannot be defined for phantom evaluations.

CNR and activity concentration ratios were evaluated as a function of the acquisition time. According to the Rose criterion, spheres with CNR ≥ 5 were considered as visible. The Rose criterion was originally derived from quantum effects in the visual process [27] and is commonly applied to define visible objects in PET imaging [26, 28]. An activity concentration percentage deviation range of ± 20% was considered acceptable based on published data on the test–retest reliability of 18F-FDG PET data [29, 30].

Resolution phantom

The system spatial resolution was determined as previously described [24] for the purpose of evaluating the image reconstruction parameters. In brief, the spatial resolution was estimated at three transversal positions (at the centre and at one-fourth of the PET scanner’s field-of-view in each direction). Gaussian functions were fitted to radial and tangential activity concentration profiles (that were determined at 10-mm and 100-mm distance from the centre at each transversal position) to estimate the respective full width at half maximum (FWHM). The system spatial resolution (FWHMsys) was defined as mean of all estimated FWHM values.

The clinical spatial resolution FWHMclin was determined from the estimated system resolution FWHMsys and the respective Gaussian filter size (FWHMGauss). The effective FWHM of the convolution of the two Gaussian functions can be calculated using the following equation [24]:

$$\begin{array}{c}{\mathrm{FWHM}}_{\mathrm{clin}} = \sqrt{{\mathrm{FWHM}}_{\mathrm{sys}}^{2} + {\mathrm{FWHM}}_{\mathrm{Gauss}}^{2}}\end{array}$$
(4)

Spatial resolution determination was performed for a voxel size of 0.83 × 0.83 x 2.00 mm3 to achieve the highest possible resolution.

Patient data evaluation

To validate the phantom results, previously published patient data [6] were re-evaluated. In brief, 18F-FDG PET data of 20 lymphoma patients were acquired on a Biograph Vision 600 PET/CT system separately at a clinical standard and at an approximately three-fold reduced total PET acquisition time. Image reconstruction was performed using OSEM-TOF and OSEM-TOF+PSF, respectively, a voxel size of 3.30 × 3.30 x 3.00 mm3 and a 4-mm Gaussian filter. For a total of 30 lesions, SUVmax and SUVpeak values were determined at standard and reduced acquisition times. These data were subjected to extended statistical analyses. Moreover, maximum intensity projection (MIP) PET images of a single patient example are presented.

Software/statistics

PET image analysis and VOI segmentation was performed using PMOD 4.202 (PMOD Technologies, Zurich, Switzerland). All statistical computations were performed using R 4.0.3 (R Foundation for Statistical Computing, Vienna, Austria, www.R-project.org). To describe the correlation between short and full acquisition time SUVs in a linear regression model, the Pearson correlation coefficient (PCC) was calculated. The two-way mixed effect model intraclass correlation coefficient (ICC) was used to describe the intra-test correlation [31, 32]; lower and upper confidence bounds were determined as defined by Shrout and Fleiss [31]. Following Koo et al. (28), an ICC ≥ 0.90 indicates excellent correlation. Bland–Altman analyses [33] and Krippendorff’s alpha [34] were used to evaluate the intra-test agreement of SUVs.

Results

Soft-tissue tumour phantom

A visual examination of the images revealed a good agreement with the CNR threshold value of CNR ≥ 5. Figure 2 exemplarily shows PET images of the soft-tissue tumour phantom for different acquisition times and image reconstruction algorithms.

Fig. 2
figure 2

Exemplary collection of PET images of the soft-tissue tumour phantom for different acquisition times and reconstruction algorithms. All images were reconstructed with a voxel size of 3.30 × 3.30 x 3.00 mm3, a 2-mm Gaussian filter, and the smallest investigated number of iterations (10i for non-TOF and 4i for TOF-based image reconstructions). The diameters of the spherical phantom inserts were 6.5 mm, 9.7 mm, 12.6 mm, 15.1 mm, 22.2 mm, and 28.0 mm (clockwise starting from bottom-left)

Contrast-to-noise ratio

For sphere diameters > 10 mm, the CNR was ≥ 5 for all investigated acquisition times, reconstruction algorithms and number of iterations. For the 6.5-mm and the 9.7-mm sphere, the detectability was dependant on the acquisition conditions (Fig. 3, CNR curves not shown for sphere diameters > 10 mm). The smallest lesion in the evaluation of patient images that was performed to select the phantom conditions was 0.7 ml. Assuming a spherical lesion, this results in a corresponding sphere diameter of 11 mm. Hence, the following evaluation is mainly performed for the 9.7-mm sphere that best matches the minimal size of patient lesions.

Fig. 3
figure 3

CNR of the 9.7-mm (A) and 6.5-mm (B) diameter soft-tissue tumour phantom spheres as a function of the acquisition time for all investigated image reconstruction algorithms and numbers of iterations (3.30 × 3.30 x 3.00-mm3 voxel size and 2-mm Gaussian filter). The sphere activity concentration was 15 kBq/mL and the SBR was 5:1 (these parameters were selected to represent clinical data). The dashed horizontal line indicates a CNR threshold value of 5

For both the 6.5- and the 9.7-mm sphere and for all acquisition conditions, the highest CNR values were observed for the lowest numbers of iterations, i.e., n = 10 for non-TOF and n = 4 for TOF-based reconstruction algorithms. OSEM-TOF+PSF performed best, followed by OSEM-TOF, OSEM-PSF, and OSEM in descending order applying a standard voxel size of 3.30 × 3.30 x 3.00 mm3 and Gaussian filter of 2 mm (Fig. 3). Therefore, the detailed evaluation was only performed for TOF-based reconstructions (with lowest number of iterations of 4): For the 9.7-mm sphere, the CNR was ≥ 5 for all evaluated acquisition times (Fig. 3A). Of note, the smallest (6.5-mm) sphere was only detectable at larger acquisition times of 210 s (OSEM-TOF, 4 iterations) and 150 s (OSEM-TOF+PSF, 4 iterations), respectively (Fig. 3B).

Varying voxel size and Gaussian filter, best results were achieved for a voxel size of 3.30 × 3.30 x 3.00 mm3 and a 4-mm Gaussian filter (Fig. 4). For a 30-s to 90-s acquisition time, the detectability criterion was met for all voxel sizes and Gaussian filters except for OSEM-TOF using a small voxel size of 1.65 × 1.65 x 2.00-mm3 and a low Gaussian smoothing level of 2 mm. For larger acquisition times, the CNR was ≥ 5 for all acquisition conditions.

Fig. 4
figure 4

CNR of the 9.7-mm diameter soft-tissue tumour phantom sphere as a function of the acquisition time for the different investigated combinations of voxel sizes and Gaussian filters, separately for OSEM-TOF (A) and OSEM-TOF+PSF (B) image reconstructions. The dashed horizontal line indicates a CNR threshold value of 5

Quantification

For the soft-tissue tumour phantom, best maximum activity concentration quantification results were achieved for a voxel size of 3.30 × 3.30 x 3.00 mm3 and a 4-mm Gaussian filter, followed by a voxel size of 1.65 × 1.65 x 2.00 mm3 and a 4-mm Gaussian filter (Fig. 5C and D for OSEM-TOF and Fig. 5G and H for OSEM-TOF+PSF). For both OSEM-TOF+PSF and OSEM-TOF (3.30 × 3.30 x 3.00-mm3 or 1.65 × 1.65 x 2.00-mm3 voxel size and 4-mm Gaussian filter) the maximum activity concentration quantification acceptance criterion was met for a 30-s acquisition time. For a 2-mm Gaussian filter (Fig. 5A and B for OSEM-TOF and Fig. 5E and F for OSEM-TOF+PSF), acquisition times of 60 to 120 s were necessary to fulfil the acceptance criterion depending on voxel size and reconstruction algorithm. The evaluation of the peak activity concentration quantification accuracy showed an acceptable quantification accuracy for a 30-s acquisition time for all examined reconstruction parameters (Supplemental Figure S1).

Fig. 5
figure 5

Maximum activity concentration ratio for the small-tumour phantom as a function of the acquisition time by reference to the 10-min acquisition time PET images separately for the different investigated combinations of voxel sizes, Gaussian filters, and OSEM-TOF (A-D) or OSEM-TOF+PSF (EH) image reconstructions. Data for all investigated spheres are presented. Dashed horizontal lines indicate the ± 20% deviation (acceptance criterion)

Bone-lung Tumour Phantom

For the bone-lung tumour phantom, best results were achieved for a voxel size of 3.30 × 3.30 x 3.00 mm3 and a 4-mm Gaussian filter (Fig. 6D and H), followed by a voxel size of 1.65 × 1.65 x 2.00 mm3 and a 4-mm Gaussian filter (Fig. 6C and G). For OSEM-TOF+PSF (3.30 × 3.30 x 3.00-mm3 voxel size and 4-mm Gaussian filter) the maximum activity concentration quantification acceptance criterion was met for a 30-s acquisition time, whereas for OSEM-TOF a 60-s acquisition time was required for accurate quantification in all three density regions. For a 2-mm Gaussian filter and a voxel size of 1.65 × 1.65 x 2.00 mm3 (Fig. 6A and E), acquisition times of 90 s (OSEM-TOF+PSF) and 150 s (OSEM-TOF) were necessary to fulfil the acceptance criterion. When increasing the voxel size to 3.30 × 3.30 x 3.00 mm3 while maintaining the 2-mm Gaussian filter (Fig. 6B and F), acquisition times of 30 s (OSEM-TOF+PSF) and 60 s (OSEM-TOF) were necessary. Similar results were achieved for the evaluation of the peak activity concentration quantification accuracy (Supplemental Figure S2). Of note, to meet the peak activity concentration acceptance criterion, even for the best performing reconstruction parameters (OSEM-TOF+PSF, 3.30 × 3.30 x 3.00 mm3-voxel size and 4-mm Gaussian filter) a 60-s acquisition time was necessary.

Fig. 6
figure 6

Maximum activity concentration ratio for the bone-lung phantom as a function of the acquisition time by reference to the 10-min acquisition time PET images separately for the different investigated combinations of voxel sizes, Gaussian filters, and OSEM-TOF (A-D) or OSEM-TOF+PSF (EH) image reconstructions. Data for all investigated spheres/density regions are presented. Dashed horizontal lines indicate the ± 20% deviation (acceptance criterion)

Resolution phantom

System spatial resolution (average FWHMsys) was 3.6 ± 0.3 mm for OSEM-TOF and 2.9 ± 0.2 mm for OSEM-TOF+PSF. Clinical spatial resolutions (FWHMclin) were 4.0 mm (OSEM-TOF) and 3.5 mm (OSEM-TOF+PSF) for a 2-mm Gaussian filter and 5.4 mm (OSEM-TOF) and 4.9 mm (OSEM-TOF+PSF) for a 4-mm Gaussian filter. All resolution phantom results are listed in Table 2.

Table 2 Detailed system spatial resolution results for the resolution phantom. All spatial resolutions were derived for a voxel size of 0.83 × 0.83 x 2.00 mm3

Evaluation of patient data

To evaluate the phantom results in a clinical context, data from a previously published patient study [6] were re-evaluated. Quantification results were compared between clinical standard and approximately three-fold reduced acquisition time PET images of lymphoma patients. For an inter-patient comparability, clinically established SUVmax and SUVpeak were compared instead of absolute activity concentration values. Overall, agreement and correlation between short and full acquisition time SUVmax values were excellent for both OSEM-TOF and OSEM-TOF+PSF reconstructed images (Fig. 7). Detailed agreement and correlation analyses results (indicating PCC, ICC, Krippendorff’s alpha and Bland–Altman bias) are presented in Table 3. Likewise, agreement and correlation of SUVpeaks were excellent for both OSEM-TOF and OSEM-TOF+PSF reconstructed images (Supplemental Figure S3 and Table 3). To indicate visual differences between short and full acquisition time PET images, a patient example is presented in Fig. 8.

Fig. 7
figure 7

Analysis of correlation and agreement of patient data (n = 31 lesions). Scatter plot (A) and Bland–Altman analysis (B) for SUVmax of reduced acquisition time versus SUVmax of full acquisition time PET images for OSEM-TOF reconstructions. Scatter plot (C) and Bland–Altman analysis (D) for SUVmax of reduced acquisition time versus SUVmax of full acquisition time PET images for OSEM-TOF+PSF image reconstructions

Table 3 Detailed agreement and correlation analysis results for evaluated patient lesions. Standard and reduced acquisition time SUVmax and SUVpeak  are compared
Fig. 8
figure 8

MIP (A&C) and transversal slice (B&D) PET images of a Non-Hodgkin Lymphoma patient who underwent FDG PET/CT for re-staging after chemotherapy. Two retroperitoneal lymphoma manifestations were detected in both reduced (orange) and full (green) acquisition time PET images. For the cranial lesion (solid arrows), SUVmax was 31.7 (34.6) for the reduced and 29.2 (31.8) for the full acquisition time images using OSEM-TOF (OSEM-TOF+PSF) image reconstruction. For the caudal lesion (dashed arrows), SUVmax was 32.7 (36.2) for the reduced and 31.3 (34.7) for the full acquisition time images using OSEM-TOF (OSEM-TOF+PSF) image reconstruction. Of note, differences in kidney activity distribution (urine activity) are caused by slightly different time-points of PET imaging

Discussion

The phantom study results indicate that the acquisition time could be reduced to 60 s per bed position (or 2.2 mm/s in continuous-bed-motion mode [35]) for 18F-FDG-PET/CT scans on the SiPM-based Biograph Vision while maintaining lesion detectability and accurate image quantification within acceptable limits. This is an approximately threefold reduction in acquisition time compared to commonly applied and recommended PET acquisition protocols [7, 16].

Several clinical evaluations for 18F-FDG and 68 Ga-PSMA suggest that an acquisition time reduction by a factor of one-third is feasible using an SiPM-based PET/CT system [6, 7]. In an evaluation by Alberts et al. [34], all lesions were detected at 50%-reduced acquisition time images, whereas 2/100 lesion were missed in 75%-reduced acquisition time images. However, a literature review revealed no published phantom study under clinically derived conditions to analyse the minimum possible acquisition time for oncologic 18F-FDG PET imaging.

Previously published phantom studies investigating SiPM-based PET/CT systems mostly use NEMA phantoms for system characterisation [2, 36,37,38,39,40] and do not investigate effects of a reduced acquisition time. One study by Gnesin et al. [36] reports a 40% to 70%-reduced time-activity-product for SiPM based PET systems indicating the possibility to reduce administered activity or acquisition time. A more sophisticated phantom study was performed by Surti et al. [41], who use a torso phantom and a cylindrical phantom to simulate lung and liver regions; lesions were scanned in air and later virtually embedded into the phantom data. They used the area under the localized receiver-operating-characteristic curve as metric to show that, compared to the PMT-based Biograph mCT, the imaging time could be reduced by a factor of 4–6 on the Vision [41]. However, no analysis of different reconstruction algorithms, number of iterations, voxel sizes, and Gaussian filters was performed in that study [41].

In this study, we used different phantoms and acquisition conditions derived from clinical lymphoma patient data to investigate a reduced acquisition time and select most beneficial reconstruction parameters. First, the CNR was investigated among different reconstruction algorithms and numbers of iterations at a standard voxel size of 3.30 × 3.30 x 3.00 mm3 and standard Gaussian smoothing level of 2 mm (Fig. 3). These evaluations were performed for the soft-tissue tumour phantom resembling nodal lymphoma lesions. TOF-based outperformed non-TOF image reconstruction algorithms and were selected for further investigation. TOF-based image reconstruction benefits in particular from the high coincidence timing resolution of SiPM-based PET systems [42]. Best results were achieved for the lowest number of iterations of 4 (in agreement with the manufacturer’s recommendation). Additional iterations led to reduced CNR values indicating increased noise; thus, convergence at 4 iterations is suggested [43]. Fast convergence of the iteration process was previously described for TOF-based image reconstruction on SiPM-based systems [44]. Of note, the further evaluations were based on the 9.7-mm diameter sphere that mimics the lower size limit of lesions in the previously performed patient evaluation and, therefore, represents the most challenging clinically realistic conditions. Likewise, the size of nodal lymphoma lesions was described as typically ≥ 10 mm [45].

Next, voxel size and level of the applied Gaussian filter were varied. Regarding detectability, small filter levels result in higher spatial resolution facilitating the detection of small lesions [46]. Regarding lesion quantification, Gaussian filtering can be used, for instance, to suppress Gibbs edge artefacts [47]. Best CNR values (for the representative 9.7-mm diameter sphere) were achieved for a 3.30 × 3.30 x 3.00-mm3 voxel size and a 4-mm Gaussian filter, followed by a 1.65 × 1.65 x 2.00-mm3 voxel size and a 4-mm Gaussian filter (Fig. 4) indicating an adequative detectability for a lesion size at the lower size range of typical clinical lesions. Of note, the Rose criterion (CNR ≥ 5) was met for all combinations and acquisition times except for 30-s to 90-s acquisition times at a 1.65 × 1.65 x 2.00-mm3 voxel size and a 2-mm Gaussian filter, and non-PSF image reconstruction.

An investigation of the quantification performance revealed that a 2-mm Gaussian filter smoothing resulted in inadequate quantification, particularly for larger lesions, at short acquisition times (Fig. 5). TOF-based image reconstruction and PSF-modelling can overcompensate for very low count statistics and lead to an overestimation of activity concentration values [48, 49]. To simulate typical extra-nodal lymphoma lesions, the evaluation was extended to the bone-lung tumour phantom and comparable results were observed (Fig. 6). Differences in CT densities compared to soft tissue can influence the scatter contribution. This might affect the PET scanner’s scatter correction that is based on tail-fitting to the PET data sinograms and, consequently, the lesion quantification performance.

Applying a Gaussian filter of 4 mm, at 60-s acquisition time satisfactory results were achieved for all phantom inserts at both 1.65 × 1.65 x 2.00-mm3 and 3.30 × 3.30 x 3.00-mm3 voxel sizes with favourable quantification performance for the larger voxel size. At 30-s acquisition time, quantification of maximum and/or peak activity concentration of the small lung-phantom insert in 1000-HU background was unsatisfactory, even for the best performing reconstruction parameters. Increasing the filter level from 2 to 4 mm resulted in a degradation of spatial resolution for OSEM-TOF (OSEM-TOF+PSF) from 4.0 mm (3.5 mm) to 5.4 mm (4.9 mm). Following the dependency of the system resolution on the sampling frequency (Huang criterion) [50], the spatial resolution could have been improved by decreasing the voxel size to 1.65 × 1.65 x 2.00-mm3. However, for the smaller voxel sizes the quantification accuracy was compromised, particularly for smaller acquisition times (Figs. 5 and 6). Moreover, the detectability of the 9.7-mm sphere was limited (CNR<5) for a 1.65 × 1.65 x 2.00-mm3 voxel size, a 2-mm Gaussian filter, and non-PSF image reconstruction as described above.

An acquisition time of 60 s per bed position signifies a threefold decrease to a clinically established acquisition time of 180 s per bed position [7]. A re-evaluation of previously acquired lymphoma patient PET data (acquired on the Biograph Vision at standard and 2.75-fold reduced acquisition times) was performed to infer on the applicability of the phantom results in a clinical setting. A detailed statistical evaluation of lesion quantification revealed an excellent correlation and agreement for SUVmax and SUVpeak between full and reduced acquisition time images for both OSEM-TOF and OSEM-TOF+PSF reconstructed images (Fig. 7, Supplemental Figure S3, and Table 3). Outliers in the Bland–Altman plots were observed for large SUVs (Fig. 7 and Supplemental Figure S3). They arise, as in the analysis a large range of SUVs (more than one magnitude) was covered. Thus, for larger SUVs, the absolute difference in SUV can be larger than for lower SUVs, even if the relative difference is low and they present as outliers in the Bland–Altman plot. The maximum percentage deviations were 18.5% (SUVmax OSEM-TOF), 17.5% (SUVmax OSEM-TOF+PSF), 14.1% (SUVpeak OSEM-TOF), and 13.7% (SUVmax OSEM-TOF+PSF), respectively. Hence, all deviations were below the 20% threshold for repeatability of 18F-FDG PET SUV measurements that was derived in an metanalysis of test–retest variability studies [29].

In summary, our data indicate that using an SiPM-based Biograph Vision PET/CT system an acquisition time of 60 s per bed position is sufficient to fulfill the lesion detectability and quantification acceptance criteria, if OSEM-TOF or OSEM-TOF+PSF image reconstruction, a 4-mm Gaussian filter, and a 1.65 × 1.65 x 2.00-mm3 or 3.30 × 3.30 x 3.00-mm3 voxel size are applied. For a further reduction in acquisition time to 30 s, not all acceptance criteria for all phantoms are met. This is in line with a published clinical evaluation of 100 lesions, in which all lesions were detected at 60 s acquisition time, whereas at 30 s acquisition time single lesions were missed [34]. Moreover, comparable results indicating a threshold of 60 s per bed position were also obtained in phantom and clinical studies investigating 68Ga-PSMA-11 PET/CT for imaging of prostate cancer patients [51, 52].

Different hardware and software improvements may lead to a further decrease in examination time/administered activity for whole-body PET examinations in the future. For example, total-body PET systems can enable reduced acquisition times [53,54,55]. In addition, artificial intelligence methods can be applied for PET image reconstruction and post-reconstruction. Several approaches demonstrated the capability of deep-learning neural networks to enhance low-count PET images [10, 56,57,58]. One software solution was already approved by the FDA and is commercially available (SubtlePET, Subtle Medical, Menlo Park, CA).

As typical PET acquisition times are long, for example, in comparison to CT acquisitions, PET imaging is influenced by patient motion including respiratory motion or patient movement. The visual impact of motion artifacts can be increased for high-resolution PET (as offered by SiPM-based PET) and for small lesions [59]. Moreover, attenuation correction can be deteriorated due to misalignment between PET and CT images. To reduce motion artifacts arising from patient movements, shorter PET acquisition times as enabled by SiPM-based systems can be of particular benefit. Influence of patient motion was not investigated in this phantom study. In the patient evaluation, the correlation between quantification results in reduced- and full-acquisition time PET images was excellent (Fig. 7, Supplemental Figure S3 and Table 3) indicating a low influence of motion artifacts in the investigated set of typical clinical lesions. In future, the implementation of total-body PET systems can be of further benefit regarding patient motion as, like in CT imaging, PET images could be acquired in respiratory breaks. Moreover, motion-correction can be performed by software- or hardware-driven approaches [60, 61].

The study faces several limitations. First, the phantoms are anthropomorphic to a limited extent. In contrast to the homogenous phantom conditions, real lymphoma manifestations can be non-spherical, inhomogeneous and of variable signal-to-background ratio. Second, the bone-lung tumour phantom represents homogeneous non-radioactive background which also deviates from real conditions. Third, the evaluation of patient data was only performed for a single choice of Gaussian filter and voxel size.

Conclusions

Based on the phantom results, an acquisition time of 60 s per bed position yields acceptable detectability and quantification results for OSEM-TOF or OSEM-TOF+PSF image reconstruction using a 4-mm Gaussian filter and a 1.65 × 1.65 x 2.00-mm3 or 3.30 × 3.30 x 3.00-mm3 voxel size on an SiPM-based PET/CT system. The corresponding threefold reduction in acquisition time coincides with previously published clinical data for FDG and PSMA PET scans. Patients might benefit from more comfortable examinations or reduced radiation exposure, if instead of reduced acquisitions times the applied activity is reduced. Larger clinical studies are warranted for further evaluation.

Availability of data and materials

The datasets generated and/or analysed during the current study are not publicly available due to privacy legislation but are available from the corresponding author on reasonable request.

Abbreviations

ACR:

Activity concentration ratio

CNR:

Contrast-to-noise ratio

EARL:

EANM Research Ltd.

FDA:

Food and Drug Administration

FWHM:

Full width at half maximum

FOV:

Field-of-view

HU:

Hounsfield units

MIP:

Maximum intensity projection

OSEM:

Ordered subset expectation maximization

PET/CT:

Positron emission tomography / Computed tomography

PET/MR:

Positron emission tomography / Magnetic resonance imaging

PSF:

Point spread function

PSMA:

Prostate specific membrane antigen

PMT:

Photomultiplier tube

SiPM:

Silicon photomultiplier

SBR:

Signal-to-background ratio

SUV:

Standardised uptake value

VOI:

Volume-of-interest

TOF:

Time of flight

References

  1. van Sluis J, Boellaard R, Somasundaram A, van Snick PH, Borra RJH, Dierckx R, Stormezand GN, Glaudemans A, Noordzij W. Image Quality and Semiquantitative Measurements on the Biograph Vision PET/CT System: Initial Experiences and Comparison with the Biograph mCT. J Nucl Med. 2020;61(1):129–35.

    Article  PubMed  CAS  Google Scholar 

  2. van Sluis J, de Jong J, Schaar J, Noordzij W, van Snick P, Dierckx R, Borra R, Willemsen A, Boellaard R. Performance Characteristics of the Digital Biograph Vision PET/CT System. J Nucl Med. 2019;60(7):1031–6.

    Article  PubMed  CAS  Google Scholar 

  3. Lopez-Mora DA, Flotats A, Fuentes-Ocampo F, Camacho V, Fernandez A, Ruiz A, Duch J, Sizova M, Domenech A, Estorch M, et al. Comparison of image quality and lesion detection between digital and analog PET/CT. Eur J Nucl Med Mol Imaging. 2019;46(6):1383–90.

    Article  PubMed  Google Scholar 

  4. Kersting D, Jentzen W, Fragoso Costa P, Sraieb M, Sandach P, Umutlu L, Conti M, Zarrad F, Rischpler C, Fendler WP, et al. Silicon-photomultiplier-based PET/CT reduces the minimum detectable activity of iodine-124. Sci Rep. 2021;11(1):17477.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  5. Kersting D, Jentzen W, Sraieb M, Costa PF, Conti M, Umutlu L, Antoch G, Nader M, Herrmann K, Fendler WP, et al. Comparing lesion detection efficacy and image quality across different PET system generations to optimize the iodine-124 PET protocol for recurrent thyroid cancer. EJNMMI Phys. 2021;8(1):14.

    Article  PubMed  PubMed Central  Google Scholar 

  6. Weber M, Jentzen W, Hofferber R, Herrmann K, Fendler WP, Rischpler C, Umutlu L, Conti M, Costa PF, Sraieb M, et al. Evaluation of (18)F-FDG PET/CT images acquired with a reduced scan time duration in lymphoma patients using the digital biograph vision. BMC Cancer. 2021;21(1):62.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  7. van Sluis J, Boellaard R, Dierckx R, Stormezand GN, Glaudemans A, Noordzij W. Image Quality and Activity Optimization in Oncologic (18)F-FDG PET Using the Digital Biograph Vision PET/CT System. J Nucl Med. 2020;61(5):764–71.

    Article  PubMed  Google Scholar 

  8. Lasnon C, Coudrais N, Houdu B, Nganoa C, Salomon T, Enilorac B, Aide N. How fast can we scan patients with modern (digital) PET/CT systems? Eur J Radiol. 2020;129: 109144.

    Article  PubMed  Google Scholar 

  9. Ansell SM. Hodgkin Lymphoma: Diagnosis and Treatment. Mayo Clin Proc. 2015;90(11):1574–83.

    Article  PubMed  Google Scholar 

  10. Chaudhari AS, Mittra E, Davidzon GA, Gulaka P, Gandhi H, Brown A, Zhang T, Srinivas S, Gong E, Zaharchuk G, et al. Low-count whole-body PET with deep learning in a multicenter and externally validated study. NPJ Digit Med. 2021;4(1):127.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Cheson BD, Fisher RI, Barrington SF, Cavalli F, Schwartz LH, Zucca E, Lister TA, Alliance AL, Lymphoma G, Eastern Cooperative Oncology G, et al. Recommendations for initial evaluation, staging, and response assessment of Hodgkin and non-Hodgkin lymphoma: the Lugano classification. J Clin Oncol. 2014;32(27):3059–68.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Uslu L, Donig J, Link M, Rosenberg J, Quon A, Daldrup-Link HE. Value of 18F-FDG PET and PET/CT for evaluation of pediatric malignancies. J Nucl Med. 2015;56(2):274–86.

    Article  PubMed  CAS  Google Scholar 

  13. Depas G, De Barsy C, Jerusalem G, Hoyoux C, Dresse MF, Fassotte MF, Paquet N, Foidart J, Rigo P, Hustinx R. 18F-FDG PET in children with lymphomas. Eur J Nucl Med Mol Imaging. 2005;32(1):31–8.

    Article  PubMed  Google Scholar 

  14. Rauscher I, Fendler WP, Hope TA, Quon A, Nekolla SG, Calais J, Richter A, Haller B, Herrmann K, Weber WA, et al. Can the Injected Dose Be Reduced in (68)Ga-PSMA-11 PET/CT While Maintaining High Image Quality for Lesion Detection? J Nucl Med. 2020;61(2):189–93.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  15. Shah DJ, Sachs RK, Wilson DJ. Radiation-induced cancer: a modern view. Br J Radiol. 2012;85(1020):e1166-1173.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  16. Boellaard R, Delgado-Bolton R, Oyen WJ, Giammarile F, Tatsch K, Eschner W, Verzijlbergen FJ, Barrington SF, Pike LC, Weber WA, et al. FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imaging. 2015;42(2):328–54.

    Article  CAS  PubMed  Google Scholar 

  17. Paes FM, Kalkanis DG, Sideras PA, Serafini AN. FDG PET/CT of extranodal involvement in non-Hodgkin lymphoma and Hodgkin disease. Radiographics. 2010;30(1):269–91.

    Article  PubMed  Google Scholar 

  18. Schaefer NG, Strobel K, Taverna C, Hany TF. Bone involvement in patients with lymphoma: the role of FDG-PET/CT. Eur J Nucl Med Mol Imaging. 2007;34(1):60–7.

    Article  PubMed  Google Scholar 

  19. Sakai O, Curtin HD, Romo LV, Som PM. Lymph node pathology: benign proliferative, lymphoma, and metastatic disease. Radiol Clin North Am. 2000;38(5):979–98.

    Article  CAS  PubMed  Google Scholar 

  20. Seltzer S: Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients, NIST Standard Reference Database 126. In.: National Institute of Standards and Technology; 1995.

  21. Chuanyong B, Ling S, Da Silva AJ, Zuo Z. A generalized model for the conversion from CT numbers to linear attenuation coefficients. IEEE Trans Nucl Sci. 2003;50(5):1510–5.

    Article  Google Scholar 

  22. Seo Y, Wong KH, Sun M, Franc BL, Hawkins RA, Hasegawa BH. Correction of photon attenuation and collimator response for a body-contouring SPECT/CT imaging system. J Nucl Med. 2005;46(5):868–77.

    PubMed  Google Scholar 

  23. Schneider W, Bortfeld T, Schlegel W. Correlation between CT numbers and tissue parameters needed for Monte Carlo simulations of clinical dose distributions. Phys Med Biol. 2000;45(2):459.

    Article  CAS  PubMed  Google Scholar 

  24. Jentzen W. An Improved Iterative Thresholding Method to Delineate PET Volumes Using the Delineation-Averaged Signal Instead of the Enclosed Maximum Signal. J Nucl Med Technol. 2015;43(1):28–35.

    Article  PubMed  Google Scholar 

  25. Nema N: NU 2–2012: Performance Measurements of Positron Emission Tomographs. Rosslyn, VA: National Electrical Manufacturers Association 2012.

  26. Carlier T, Eugène T, Bodet-Milin C, Garin E, Ansquer C, Rousseau C, Ferrer L, Barbet J, Schoenahl F, Kraeber-Bodéré F. Assessment of acquisition protocols for routine imaging of Y-90 using PET/CT. EJNMMI Res. 2013;3(1):11.

    Article  PubMed  PubMed Central  Google Scholar 

  27. Rose A. Vision: human and electronic. New York: Plenum Press; 1973.

    Google Scholar 

  28. Adler S, Seidel J, Choyke P, Knopp MV, Binzel K, Zhang J, Barker C, Conant S, Maass-Moreno R. Minimum lesion detectability as a measure of PET system performance. EJNMMI Phys. 2017;4(1):13.

    Article  PubMed  PubMed Central  Google Scholar 

  29. de Langen AJ, Vincent A, Velasquez LM, van Tinteren H, Boellaard R, Shankar LK, Boers M, Smit EF, Stroobants S, Weber WA, et al. Repeatability of 18F-FDG uptake measurements in tumors: a metaanalysis. J Nucl Med. 2012;53(5):701–8.

    Article  PubMed  CAS  Google Scholar 

  30. Kurland BF, Peterson LM, Shields AT, Lee JH, Byrd DW, Novakova-Jiresova A, Muzi M, Specht JM, Mankoff DA, Linden HM, et al. Test-Retest Reproducibility of (18)F-FDG PET/CT Uptake in Cancer Patients Within a Qualified and Calibrated Local Network. J Nucl Med. 2019;60(5):608–14.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  31. Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979;86(2):420–8.

    Article  CAS  PubMed  Google Scholar 

  32. Koo TK, Li MY. A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research. J Chiropr Med. 2016;15(2):155–63.

    Article  PubMed  PubMed Central  Google Scholar 

  33. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–10.

    Article  CAS  PubMed  Google Scholar 

  34. Alberts I, Sachpekidis C, Prenosil G, Viscione M, Bohn KP, Mingels C, Shi K, Ashar-Oromieh A, Rominger A. Digital PET/CT allows for shorter acquisition protocols or reduced radiopharmaceutical dose in [(18)F]-FDG PET/CT. Ann Nucl Med. 2021;35(4):485–92.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  35. Osborne DR, Acuff S, Cruise S, Syed M, Neveu M, Stuckey A, Bradley Y. Quantitative and qualitative comparison of continuous bed motion and traditional step and shoot PET/CT. Am J Nuclear Med Mol Imaging. 2015;5(1):56.

    Google Scholar 

  36. Gnesin S, Kieffer C, Zeimpekis K, Papazyan JP, Guignard R, Prior JO, Verdun FR, Lima TVM. Phantom-based image quality assessment of clinical (18)F-FDG protocols in digital PET/CT and comparison to conventional PMT-based PET/CT. EJNMMI Phys. 2020;7(1):1.

    Article  PubMed  PubMed Central  Google Scholar 

  37. Carlier T, Ferrer L, Conti M, Bodet-Milin C, Rousseau C, Bercier Y, Bendriem B, Kraeber-Bodere F. From a PMT-based to a SiPM-based PET system: a study to define matched acquisition/reconstruction parameters and NEMA performance of the Biograph Vision 450. EJNMMI Phys. 2020;7(1):55.

    Article  PubMed  PubMed Central  Google Scholar 

  38. Lindstrom E, Lindsjo L, Sundin A, Sorensen J, Lubberink M. Evaluation of block-sequential regularized expectation maximization reconstruction of (68)Ga-DOTATOC, (18)F-fluoride, and (11)C-acetate whole-body examinations acquired on a digital time-of-flight PET/CT scanner. EJNMMI Phys. 2020;7(1):40.

    Article  PubMed  PubMed Central  Google Scholar 

  39. Miwa K, Wagatsuma K, Nemoto R, Masubuchi M, Kamitaka Y, Yamao T, Hiratsuka S, Yamaguchi M, Yoshii T, Kobayashi R, Miyaji N, Ishii K. Detection of sub-centimeter lesions using digital TOF-PET/CT system combined with Bayesian penalized likelihood reconstruction algorithm. Ann Nucl Med. 2020;34(10):762-771. https://doi.org/10.1007/s12149-020-01500-8. Epub 2020 Jul 4. PMID: 32623569.

  40. Koopman D, Groot Koerkamp M, Jager PL, Arkies H, Knollema S, Slump CH, Sanches PG, van Dalen JA. Digital PET compliance to EARL accreditation specifications. EJNMMI Phys. 2017;4(1):9.

    Article  PubMed  PubMed Central  Google Scholar 

  41. Surti S, Viswanath V, Daube-Witherspoon ME, Conti M, Casey ME, Karp JS. Benefit of Improved Performance with State-of-the Art Digital PET/CT for Lesion Detection in Oncology. J Nucl Med. 2020;61(11):1684-1690. https://doi.org/10.2967/jnumed.120.242305. Epub 2020 Mar 20. PMID: 32198313.

  42. Conti M, Bendriem B. The new opportunities for high time resolution clinical TOF PET. Clin Trans Imaging. 2019;7(2):139–47.

    Article  Google Scholar 

  43. Boellaard R, Van Lingen A, Lammertsma AA. Experimental and clinical evaluation of iterative reconstruction (OSEM) in dynamic PET: quantitative characteristics and effects on kinetic modeling. J Nucl Med. 2001;42(5):808–17.

    CAS  PubMed  Google Scholar 

  44. Salvadori J, Odille F, Verger A, Olivier P, Karcher G, Marie PY, Imbert L. Head-to-head comparison between digital and analog PET of human and phantom images when optimized for maximizing the signal-to-noise ratio from small lesions. EJNMMI Phys. 2020;7(1):11.

    Article  PubMed  PubMed Central  Google Scholar 

  45. Haque W, Voong KR, Shihadeh F, Arzu I, Pinnix C, Mazloom A, Medeiros LJ, Romaguera J, Rodriguez A, Wang M, et al. Radiation therapy is an effective modality in the treatment of mantle cell lymphoma, even in heavily pretreated patients. Clin Lymphoma Myeloma Leuk. 2014;14(6):474–9.

    Article  PubMed  Google Scholar 

  46. Kadrmas DJ, Casey ME, Black NF, Hamill JJ, Panin VY, Conti M. Experimental comparison of lesion detectability for four fully-3D PET reconstruction schemes. IEEE Trans Med Imaging. 2009;28(4):523–34.

    Article  PubMed  Google Scholar 

  47. Armstrong IS, Kelly MD, Williams HA, Matthews JC. Impact of point spread function modelling and time of flight on FDG uptake measurements in lung lesions using alternative filtering strategies. EJNMMI physics. 2014;1(1):99.

    Article  PubMed  PubMed Central  Google Scholar 

  48. Ptáček J, Karhan P, Fiala P. Optimal reconstruction matrix and PET image filtration for point-spread function and time-of-flight reconstruction – A phantom study. Physica Med. 2017;39:95–9.

    Article  Google Scholar 

  49. Shekari M, Ghafarian P, Ahangari S, Ay MR. Quantification of the impact of TOF and PSF on PET images using the noise-matching concept: clinical and phantom study. Nucl Sci Tech. 2017;28(11):167.

    Article  Google Scholar 

  50. Huang S-C, Hoffman EJ, Phelps ME, Kuhl DE. Quantitation in positron emission computed tomography: 3 Effect of sampling. J Comput Assist Tomogr. 1980;4(6):819–26.

    Article  CAS  PubMed  Google Scholar 

  51. Fragoso Costa P, Jentzen W, SÜßELBECK F, Fendler WP, Rischpler C, Herrmann K, Conti M, Kersting D, Weber M. Reduction of emission time for [68Ga]Ga-PSMA PET/CT using the digital biograph vision: a Phantom study. Q J Nucl Med Mol Imaging. 2021. https://doi.org/10.23736/S1824-4785.21.03300-8. Epub ahead of print. PMID: 34309334.

  52. Weber M, Jentzen W, Hofferber R, Herrmann K, Fendler WP, Conti M, Wetter A, Kersting D, Rischpler C, Fragoso Costa P. Evaluation of [(68)Ga]Ga-PSMA PET/CT images acquired with a reduced scan time duration in prostate cancer patients using the digital biograph vision. EJNMMI Res. 2021;11(1):21.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  53. Zhang YQ, Hu PC, Wu RZ, Gu YS, Chen SG, Yu HJ, Wang XQ, Song J, Shi HC. The image quality, lesion detectability, and acquisition time of (18)F-FDG total-body PET/CT in oncological patients. Eur J Nucl Med Mol Imaging. 2020;47(11):2507–15.

    Article  PubMed  CAS  Google Scholar 

  54. Hu P, Zhang Y, Yu H, Chen S, Tan H, Qi C, Dong Y, Wang Y, Deng Z, Shi H. Total-body (18)F-FDG PET/CT scan in oncology patients: how fast could it be? Eur J Nucl Med Mol Imaging. 2021;48(8):2384–94.

    Article  PubMed  Google Scholar 

  55. Alberts I, Hunermund JN, Prenosil G, Mingels C, Bohn KP, Viscione M, Sari H, Vollnberg B, Shi K, Afshar-Oromieh A, et al. Clinical performance of long axial field of view PET/CT: a head-to-head intra-individual comparison of the Biograph Vision Quadra with the Biograph Vision PET/CT. Eur J Nucl Med Mol Imaging. 2021;48(8):2395–404.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  56. Sanaat A, Shiri I, Arabi H, Mainta I, Nkoulou R, Zaidi H. Deep learning-assisted ultra-fast/low-dose whole-body PET/CT imaging. Eur J Nucl Med Mol Imaging. 2021;48(8):2405–15.

    Article  PubMed  PubMed Central  Google Scholar 

  57. Lei Y, Dong X, Wang T, Higgins K, Liu T, Curran WJ, Mao H, Nye JA, Yang X. Whole-body PET estimation from low count statistics using cycle-consistent generative adversarial networks. Phys Med Biol. 2019;64(21):215017.

    Article  PubMed  PubMed Central  Google Scholar 

  58. Kaplan S, Zhu YM. Full-Dose PET Image Estimation from Low-Dose PET Image Using Deep Learning: a Pilot Study. J Digit Imaging. 2019;32(5):773–8.

    Article  PubMed  Google Scholar 

  59. Daou D. Respiratory motion handling is mandatory to accomplish the high-resolution PET destiny. Eur J Nucl Med Mol Imaging. 2008;35(11):1961–70.

    Article  PubMed  Google Scholar 

  60. Buther F, Jones J, Seifert R, Stegger L, Schleyer P, Schafers M. Clinical Evaluation of a Data-Driven Respiratory Gating Algorithm for Whole-Body PET with Continuous Bed Motion. J Nucl Med. 2020;61(10):1520–7.

    Article  PubMed  CAS  Google Scholar 

  61. Messerli M, Liberini V, Grunig H, Maurer A, Skawran S, Lohaus N, Husmann L, Orita E, Trinckauf J, Kaufmann PA, et al. Clinical evaluation of data-driven respiratory gating for PET/CT in an oncological cohort of 149 patients: impact on image quality and patient management. Br J Radiol. 2021;94(1126):20201350.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

None.

Funding

Open Access funding enabled and organized by Projekt DEAL. This work was supported by the Universitätsmedizin Essen Clinician Scientist Academy (UMEA)/German Research Foundation (DFG, Deutsche Forschungsgemeinschaft) under Grant FU356/12–1 to DK.

Author information

Authors and Affiliations

Authors

Contributions

P.F.C., W.J., and D.K. designed the study, co-wrote the manuscript, and approved of its final content. A.B. and W.J. acquired the data. P.F.C., W.J., A.B., M.W., M.S., and D.K. analysed the data. A.B., I.M., F.Z., L.U., W.P.F., C.R., K.H., M.C., R.S., M.S, and M.W. contributed to the study design, critically revised the manuscript, and approved of its final content. The author(s) read and approved the final manuscript.

Corresponding author

Correspondence to David Kersting.

Ethics declarations

Ethics approval and consent to participate

The retrospective analysis of available patient data was performed in accordance with the declaration of Helsinki and approved by the institutional ethics committee (University of Duisburg-Essen, medical faculty, ethics protocol number 20–9204-BO). Written informed consent was obtained by all participants.

Consent for publication

Not applicable.

Competing interests

Walter Jentzen received research funding from Siemens Healthineers. Lale Umutlu is a Speaker / Advisory Board Member for Bayer Healthcare and Siemens Healthcare and received research grants from Siemens Healthcare, outside of the submitted work. Wolfgang Peter Fendler reports fees from BTG (consultant), Calyx (consultant), RadioMedix (image reader), Bayer (speakers bureau), and Parexel (image reader) outside of the submitted work. Christoph Rischpler reports a research grant from Pfizer, consultancy for Adacap and Pfizer, speaker honoraria from Adacap, Alnylam, BTG, Curium, GE Healthcare, Pfizer and Siemens Healthineers., outside of the submitted work. Ken Herrmann reports personal fees from Bayer SIRTEX Adacap Curium Endocyte IPSEN Siemens Healthineers GE Healthcare Amgen Novartis and ymabs personal fees and other from Sofie Biosciences non-financial support from ABX grants and personal fees from BTG, outside of the submitted work. Maurizio Conti is a full-time employee of Siemens Medical Solutions USA, Inc. Manuel Weber reports fees from Boston Scientific, Terumo, Eli Lilly, and Advanced Accelerator Applications, outside of the submitted work. David Kersting reports a research grant from Pfizer, outside of the submitted work. Pedro Fragoso Costa, Alissa Brahmer, Ilektra-Antonia Mavroeidi, Fadi Zarrad, Robert Seifert, and Miriam Sraieb have nothing to declare.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1: Supplemental Figure S1.

Peak activity concentration ratio for the small-tumour phantom as a function of the acquisition time by reference to the 10-min acquisition time PET images separately for the different investigated combinations of voxel sizes, Gaussian filters, and OSEM-TOF (A-D) or OSEM-TOF+PSF (E-H) image reconstructions. Data for all investigated spheres are presented. Dashed horizontal lines indicate the ±20% deviation (acceptance criterion). Supplemental Figure S2. Peak activity concentration ratio for the bone-lung phantom as a function of the acquisition time by reference to the 10-min acquisition time PET images separately for the different investigated combinations of voxel sizes, Gaussian filters, and OSEM-TOF (A-D) or OSEM-TOF+PSF (E-H) image reconstructions. Data for all investigated spheres/density regions are presented. Dashed horizontal lines indicate the ±20% deviation (acceptance criterion). Supplemental Figure S3. Analysis of correlation and agreement of patient data. Scatter plot (A) and Bland-Altman analysis (B) for SUVpeak of reduced acquisition time versus SUVpeak of full acquisition time PET images for OSEM-TOF reconstructions. Scatter plot (C) and Bland-Altman analysis (D) for SUVpeak of reduced acquisition time versus SUVpeak of full acquisition time PET images for OSEM-TOF+PSF image reconstructions.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Fragoso Costa, P., Jentzen, W., Brahmer, A. et al. Phantom-based acquisition time and image reconstruction parameter optimisation for oncologic FDG PET/CT examinations using a digital system. BMC Cancer 22, 899 (2022). https://doi.org/10.1186/s12885-022-09993-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12885-022-09993-4

Keywords