Skip to main content

Table 3 Validation metrics of MammaTyper amplification for each different target sequence and sample type

From: Technical validation of an RT-qPCR in vitro diagnostic test system for the determination of breast cancer molecular subtypes by quantification of ERBB2, ESR1, PGR and MKI67 mRNA levels from formalin-fixed paraffin-embedded breast tumor specimens

Versant kPCR

ERBB2

ESR1

PGR

MKI67

B2M

CALM2

FFPE Master Sample

LoD [Cq]

34.36

35.54

32.33

35.66

35.80

34.78

LoD [molecules]

10

5

87

7

7

8

AMR [Cq]

21.05–34.36

19.13–35.54

21.13–32.33

23.97–35.66

16.78–35.80

23.90–34.78

Efficiency [%]

106

105

106

101

100

111

Linearity

0.999

0.999

0.995

0.999

0.999

0.998

IVT-RNA

LoD [Cq]

35.31

35.60

36.72

36.41

36.22

33.69

LoD [molecules]

5

5

5

5

5

18

AMR [Cq]

12.36–35.31

12.45–35.60

12.64–36.72

11.63–36.41

12.41–36.22

11.67–33.69

Efficiency [%]

105

104

100

99

100

100

Linearity

0.999

0.999

1.000

0.999

1.000

1.000

LightCycler 480 II

ERBB2

ESR1

PGR

MKI67

B2M

CALM2

FFPE Master Sample

LoD [Cq]

33.56

36.47

33.56

35.47

35.6

35.62

LoD [molecules]

40

8

64

12

12

9

AMR [Cq]

22.32–33.56

21.69–36.47

21.20–33.56

24.61–35.47

17.60–35.60

25.26–35.62

Efficiency [%]

109

113

115

111

109

117

Linearity

1.000

1.000

0.996

0.999

0.999

0.997

IVT-RNA

LoD [Cq]

34.64

35.21

35.44

34.91

34.97

36.51

LoD [molecules]

18

18

18

18

18

5

AMR [Cq]

17.34–34.64

18.52–35.21

17.56–35.44

19.27–34.91

17.9–34.97

18.87–36.51

Efficiency [%]

106

109

101

103

107

103

Linearity

1.000

0.999

1.000

0.999

0.999

1.000