Skip to main content

Table 1 Planning constraints of planning target volume, normal organs and planning organ at risk volumes for spine stereotactic ablative radiotherapy plans

From: Progressive resolution optimizer (PRO) predominates over photon optimizer (PO) in sparing of spinal cord for spine SABR VMAT plans

Structure

Planning constraints

Priority

PTV

D0% < 18.5 Gy

200

D100% > 18 Gy

150

Spinal cord (or spinal cord PRV)

D1.2 cc < 7 Gy

120

D0.35 cc < 10 Gy

120

D0.035 cc < 14 Gy

150

Cauda equine (or cauda equine PRV)

D5 cc < 14 Gy

120

D0.035 cc < 16 Gy

150

  1. Note: PTV = planning target volume; PRV = planning organ at risk volume;