Table 3 The deviation in the suggested activity and lung doses between standard whole body scan LSFwb (St) and models of LSFwb (SC): from scatter corrected whole-body scan, LSFspect (AC-SC): from SPECT/CT with attenuation and scatter correction, and LSFspect (NoAC-SC): from SPECT with only scatter correction

From: Simplification of dosimetry in 90Y-radioembolization therapy by dual planar images

 

LSFspect

(AC-SC)

LSFspect

(NoAC-SC)

LSFwb

(SC)

Suggested Activity

Mean − 5.3%

Mean − 3.2%

Mean − 5.5%

Range (−2/−11)

Range (−1/−9)

Range (− 2/− 7)

Pvalue = 0.55

Pvalue = 0.58

Pvalue = 0.52

Lung Dose

Mean − 53%

Mean − 33%

Mean − 48%

Range (− 34/− 72)

Range (−12/− 64)

Range (− 18/− 72)

Pvalue = 0.02

Pvalue = 0.12

Pvalue = 0.03