Skip to main content

Table 5 Sensitivity, Specificity and Positive and Negative Predictive Values of Electroimpedance Mammography EIM

From: Monofrequency electrical impedance mammography (EIM) diagnostic system in breast cancer screening

BI-RADS EIM

Benign

Malignant

4, 5

44

6

1,2,3

1149

1

Sensitivity

85%

 

Specificity

96%

 

Positive predictive value (PPV)

12%

 

Negative predictive value (NPV)

99%